FDA Adverse Event Malfunction Summary report: N

INSTINCT ENDOSCOPIC HEMOCLIP

MDR report key: 3222169 · Received July 3, 2013

Report

Report Number
1037905-2013-00505
Event Type
Malfunction
Date Received
July 3, 2013
Date of Event
May 31, 2013
Report Date
June 4, 2013
Manufacturer
WILSON-COOK MEDICAL INC
Product Code
MND
PMA / PMN Number
K121505
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION EVALUATION: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED, FOUND THE DRIVE WIRE SEPARATED INSIDE THE HANDLE. THE PROXIMAL END OF THE DEVICE WIRE APPEARED ALTERED OR DAMAGED AND DID NOT SHOW THE TYPICAL BOWING INDICATIVE OF PROPER ASSEMBLY OF THE SECURING COMPONENT. THEREFORE, THE HANDLE SPOOL WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE SECURING COMPONENT TIP. THE TIP SHOWED DEFORMATION WHERE IT CONTACTED THE DRIVE WIRE THUS INDICATING PROPER ASSEMBLY OF THE SECURING COMPONENT. USING HEMOSTATS TO GRASP THE DRIVE WIRE, THE CLIP WAS OPENED AND CLOSED. A SLIGHT INCREASE IN RESISTANCE WAS OBSERVED IN THE MOVEMENT OF THE DRIVE WIRE ONCE THE CLIP WAS PULLED HALFWAY INTO THE HOUSING PRIOR TO DEPLOYMENT. THE DEVICE WAS ADVANCED INTO OLYMPUS GIF Q20 2.8 ENDOSCOPE IN A SIMULATED UPPER GI POSITION WITH THE TIP IN THE MAXIMUM RETROFLEXED POSITION TO SIMULATE A WORST CASE POSITION. USING THE HEMOSTATS, THE CLIP DID SUCCESSFULLY DEPLOY ON SIMULATED TISSUE. THE DRIVE WIRE WAS REMOVED FROM THE DEVICE, VISUALLY EXAMINED, AND DETERMINED THAT THE DRIVE WIRE WAS MANUFACTURED CORRECTLY. THE HOOK OF THE DRIVE WIRE IN INTACT. A PRODUCT-SPECIFIC DISCREPANCY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THIS OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: RESISTANCE ENCOUNTERED WHILE ATTEMPTING TO DEPLOY THE CLIP AFTER THE CLIP IS CLOSED ON THE TISSUE (I.E. DIFFICULTY WITH CLIP RELEASE) CAN LEAD TO DAMAGE OF DEVICE COMPONENTS SUCH AS THE HOOK AT THE DISTAL END OF THE DRIVE WIRE OR THE SECURITY OF THE DRIVE WIRE TO THE HANDLE. THE INSTRUCTIONS FOR USE INSTRUCTS THE USER TO VERIFY SMOOTH HANDLE OPERATION AND CLIP OPERATION PRIOR TO USE. INSTRUCTIONS FOR USE STATES TO PERMANENTLY DEPLOY CLIP, PULL HANDLE SPOOL TOWARD HANDLE THUMB RING UNTIL CLIP DETACHES. NOTE: IF SEPARATION OF CLIP IS NOT IMMEDIATE, GENTLY MOVE CATHETER BACK AND FORTH OR USE OTHER ENDOSCOPIC MANEUVERS TO SEPARATE CATHETER FROM CLIP. PRIOR TO DISTRIBUTION, ALL INSTINCT ENDOSCOPIC HEMOCLIPS ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.

Description of Event or Problem · 1

A COOK INSTINCT ENDOSCOPIC HEMOCLIP WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD). WHEN THE PHYSICIAN CLOSED THE CLIP ONTO THE MUCOSA, THE DRIVE WIRE BROKE INSIDE THE HAND PIECE. A KELLY CLAMP WAS USED TO GRAB ONTO THE DRIVE WIRE AND THE PHYSICIAN WAS ABLE TO OPEN THE CLIP FROM THE MUCOSA AND THEN REMOVE THE CLIP FROM THE ENDOSCOPE. ANOTHER COOK DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303852 INSTINCT ENDOSCOPIC HEMOCLIP MND WILSON-COOK MEDICAL INC W3257139

Patients

Seq Age Sex Outcome Treatment
1 OLYMPUS ENDOSCOPE (UNKNOWN MODEL NUMBER)