FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3222063
·
Received July 1, 2013
Report
- Report Number
- 1314492-2013-00944
- Event Type
- Malfunction
- Date Received
- July 1, 2013
- Date of Event
- June 3, 2013
- Report Date
- June 3, 2013
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. EVALUATION COULD NOT REPRODUCE THE REPORTED SYMPTOM. A DEVICE HISTORY LOG REVIEW WAS CONDUCTED AND THERE WERE NO OCCURRENCES OF SYSTEM ERROR 322. THE DEVICE WAS DETERMINED TO BE WITHIN SPECIFICATION IN RELATION TO THE REPORTED SYMPTOM. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PUMP ALARMED FOR A SYSTEM ERROR 322 - "LNK SWITCH ERROR (LOW)." IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 298256 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |