FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3221963 · Received July 5, 2013

Report

Report Number
1314492-2013-01022
Event Type
Malfunction
Date Received
July 5, 2013
Date of Event
June 1, 2013
Report Date
June 7, 2013
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED FOR EVALUATION AND BAXTER WAS ABLE TO CONFIRM THAT THE AUDIO FUNCTION WAS NOT PRESENT. FURTHER EVALUATION IDENTIFIED THAT THE ABSENCE OF AUDIO WAS DUE TO A FAILED SPEAKER. ALL DETECTION CAPABILITIES AND FUNCTIONS PERFORMED AS EXPECTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP DID NOT HAVE AUDIO FUNCTION. IT WAS ALSO REPORTED THAT THIS WAS FOUND DURING TESTING AND THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
307115 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1