SPECTRUM INFUSION PUMP
MDR report key: 3221936
·
Received July 5, 2013
Report
- Report Number
- 1314492-2013-01014
- Event Type
- Malfunction
- Date Received
- July 5, 2013
- Date of Event
- June 1, 2013
- Report Date
- June 6, 2013
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE IN QUESTION. THE EVALUATION CONFIRMED THE LOWER READING ON THE ULTRASONIC SENSOR TO BE BELOW SPECIFICATION CAUSED BY A FAILED UPSTREAM SENSOR. WITH LOW ULTRASONIC READINGS, THE DEVICE WILL BECOME MORE SENSITIVE TO AIR AND UPSTREAM OCCLUSION ALARMS. THE FAILED UPSTREAM SENSOR WAS REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PUMP ALARMS FREQUENTLY FOR UPSTREAM OCCLUSIONS (MEDICATION, PROGRAMMED AMOUNT AND DELIVERY RATE UNKNOWN). THE CUSTOMER STATED THAT THERE WAS NO PATIENT INJURY AND NO ADDITIONAL INFORMATION COULD BE PROVIDED.