TERUMO ADVANCED PERFUSION SYSTEM 1
Report
- Report Number
- 1828100-2013-00645
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Date of Event
- June 17, 2013
- Report Date
- June 18, 2013
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEM CORP
- Product Code
- DTQ
- PMA / PMN Number
- K022947
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE FIELD SERVICE REP (FSR) WAS UNABLE TO VERIFY STATED ISSUE. THE PERFUSIONIST (CCP) MENTIONED HE HAD PERFORMED A SCREEN CALIBRATION AFTER THE ISSUE AND PREVIOUSLY BEFORE THIS INCIDENT BUT THE MONITOR WOULD COME OUT OF CALIBRATION. THE FSR INSTALLED AND CONFIGURED A SERVICE LOANER MONITOR, AND PERFORMED FUNCTIONAL TESTING. THE LOANER UNIT OPERATED TO MANUFACTURER SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. THE SUSPECT DEVICE WAS RETURNED TO THE MANUFACTURER FOR FURTHER EVAL.
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE TOUCH SCREEN WAS NOT RESPONSIVE ON THE CENTRAL CONTROL MONITOR (CCM). THE CURSOR WAS ALSO OUT OF ALIGNMENT (IT WAS AROUND AN INCH HIGHER THAN WHERE CUSTOMER WANTED TO SELECT ON THE SCREEN). AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO DELAYS, NO BLOOD LOSS OR NO ADVERSE CONSEQUENCES TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312882 | TERUMO ADVANCED PERFUSION SYSTEM 1 | DTQ | TERUMO CARDIOVASCULAR SYSTEM CORP | 802100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |