TERUMO TCM II COOLING AND HEATING SYSTEM
Report
- Report Number
- 1828100-2013-00650
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Date of Event
- June 18, 2013
- Report Date
- June 18, 2013
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEM CORP
- Product Code
- DWC
- PMA / PMN Number
- K883603
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE BIOMED MEASURED THE FLOW IN "LARGE TANK MODE" AND IT MEASURED AT 4.2 LITERS PER MINUTE AT HIGH SPEED. HE SAID IN "LARGE TO SMALL" MODE, IT MEASURED 17 LITERS PER MINUTE. THE BIOMED ASKED TECHNICAL SUPPORT IF THE HANSEN FITTINGS COULD MAKE THAT MUCH OF A DIFFERENCE. TECHNICAL SUPPORT TOLD THE BIOMED, THAT IF THEY ARE PARTIALLY CLOGGED IT WOULD AFFECT THE FLOW. THE TRAINED BIOMED REPLACED THE HANSEN FITTINGS AND THE FLOW ON THE COOLER HEATER UNIT WAS NOW NORMAL. THE OLD FITTINGS WERE DISCARDED AFTER BEING REPLACED, AND NO PARTS CAN BE RETURNED TO THE MANUFACTURER FOR ANALYSIS. IF ADDITIONAL INFO BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
THE USER FACILITY'S BIOMEDICAL ENGINEER (BIOMED) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE CARDIOPLEGIA WAS NOT FLOWING AS IT WOULD NORMALLY BE IN "LARGE TANK MODE". SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312835 | TERUMO TCM II COOLING AND HEATING SYSTEM | DWC | TERUMO CARDIOVASCULAR SYSTEM CORP | 4415 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |