LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM WITH RAPIDPORT
Report
- Report Number
- 2024601-2013-00577
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Date of Event
- January 31, 2013
- Report Date
- June 13, 2013
- Manufacturer
- ALLERGAN
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE REPORTER OF THE COMPLAINT WAS ASKED TO RETURN THE PRODUCT FOR ANALYSIS IF IT IS EXPLANTED IN THE FUTURE. THE DEVICE WAS RESUTURED INTO PLACE AND REMAINS IMPLANTED. BASED UPON THE MODEL NUMBER AND IMPLANT DATE PROVIDED BY THE REPORTER THE CONNECTOR TYPE IS ASSUMED TO BE A RAPIDPORT EZ STRAIN RELIEF. VISUAL EXAMINATION MAY DETERMINE THE CONNECTOR TYPE ASSOCIATED WITH THIS REPORT. THE REPORTER OF THE EVENT WAS ASKED TO RETURN THE PRODUCT FOR ANALYSIS, IF IT IS EXPLANTED IN THE FUTURE. IT WAS REPOSITIONED DURING THE PROCEDURE AND ALLERGAN HAS NOT RECEIVED THE PRODUCT AT THIS TIME. THEREFORE NO ANALYSIS OR TESTING HAS BEEN DONE. NO ADDITIONAL INFORMATION HAS BEEN REPORTED TO ALLERGAN REGARDING THE SERIAL NUMBER. DEVICE LABELING ADDRESSES THE REPORTED EVENT OF DISPLACEMENT AS FOLLOWS: "CAUTION: "CARE MUST BE TAKEN TO PLACE THE ACCESS PORT IN A STABLE POSITION AWAY FROM AREAS THAT MAY BE AFFECTED BY SIGNIFICANT WEIGHT LOSS, PHYSICAL ACTIVITY, OR SUBSEQUENT SURGERY. FAILURE TO DO SO MAY RESULT IN THE INABILITY TO PERFORM PERCUTANEOUS BAND ADJUSTMENTS." "ACCESS PORTS HAVE BEEN REPORTED TO BE "FLIPPED" OR INVERTED. IF YOU INITIALLY SEE AN OBLIQUE OR SIDE VIEW ON X-RAY THEN EITHER REPOSITION THE PATIENT OR THE X-RAY EQUIPMENT UNTIL YOU OBTAIN A PERPENDICULAR, OVERHEAD (0 DEGREE) VIEW. TARGETING THE PORT FOR NEEDLE PENETRATION CAN BE DIFFICULT IF THIS ORIENTATION IS NOT CONTROLLED." "CAUTION: THE ACCESS PORT MUST BE SECURELY FASTENED TO THE PATIENT'S RECTUS FASCIA WITH ALL FOUR OF THE STAINLESS STEEL ANCHORS FIRMLY EMBEDDED IN THE PATIENT'S FASCIA. IF THIS IS NOT ACHIEVED, THE ACCESS PORT MAY BECOME LOOSE AND INACCESSIBLE FOR POST-OPERATIVE ADJUSTMENTS, THUS REQUIRING REVISIONAL SURGERY."
HEALTHCARE PROFESSIONAL REPORTED, RAPIDPORT EZ ACCESS PORT FLIPPED 180 DEGREES." "COULD NOT ACCESS RAPIDPORT EZ ACCESS PORT DURING ROUTINE CLINIC VISIT, PATIENT SENT TO X-RAY AND PORT FLIP IDENTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312832 | LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM WITH RAPIDPORT | LTI | ALLERGAN | NA | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR |