FDA Adverse Event Malfunction Summary report: N

LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM WITH RAPIDPORT

MDR report key: 3221835 · Received July 9, 2013

Report

Report Number
2024601-2013-00577
Event Type
Malfunction
Date Received
July 9, 2013
Date of Event
January 31, 2013
Report Date
June 13, 2013
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTER OF THE COMPLAINT WAS ASKED TO RETURN THE PRODUCT FOR ANALYSIS IF IT IS EXPLANTED IN THE FUTURE. THE DEVICE WAS RESUTURED INTO PLACE AND REMAINS IMPLANTED. BASED UPON THE MODEL NUMBER AND IMPLANT DATE PROVIDED BY THE REPORTER THE CONNECTOR TYPE IS ASSUMED TO BE A RAPIDPORT EZ STRAIN RELIEF. VISUAL EXAMINATION MAY DETERMINE THE CONNECTOR TYPE ASSOCIATED WITH THIS REPORT. THE REPORTER OF THE EVENT WAS ASKED TO RETURN THE PRODUCT FOR ANALYSIS, IF IT IS EXPLANTED IN THE FUTURE. IT WAS REPOSITIONED DURING THE PROCEDURE AND ALLERGAN HAS NOT RECEIVED THE PRODUCT AT THIS TIME. THEREFORE NO ANALYSIS OR TESTING HAS BEEN DONE. NO ADDITIONAL INFORMATION HAS BEEN REPORTED TO ALLERGAN REGARDING THE SERIAL NUMBER. DEVICE LABELING ADDRESSES THE REPORTED EVENT OF DISPLACEMENT AS FOLLOWS: "CAUTION: "CARE MUST BE TAKEN TO PLACE THE ACCESS PORT IN A STABLE POSITION AWAY FROM AREAS THAT MAY BE AFFECTED BY SIGNIFICANT WEIGHT LOSS, PHYSICAL ACTIVITY, OR SUBSEQUENT SURGERY. FAILURE TO DO SO MAY RESULT IN THE INABILITY TO PERFORM PERCUTANEOUS BAND ADJUSTMENTS." "ACCESS PORTS HAVE BEEN REPORTED TO BE "FLIPPED" OR INVERTED. IF YOU INITIALLY SEE AN OBLIQUE OR SIDE VIEW ON X-RAY THEN EITHER REPOSITION THE PATIENT OR THE X-RAY EQUIPMENT UNTIL YOU OBTAIN A PERPENDICULAR, OVERHEAD (0 DEGREE) VIEW. TARGETING THE PORT FOR NEEDLE PENETRATION CAN BE DIFFICULT IF THIS ORIENTATION IS NOT CONTROLLED." "CAUTION: THE ACCESS PORT MUST BE SECURELY FASTENED TO THE PATIENT'S RECTUS FASCIA WITH ALL FOUR OF THE STAINLESS STEEL ANCHORS FIRMLY EMBEDDED IN THE PATIENT'S FASCIA. IF THIS IS NOT ACHIEVED, THE ACCESS PORT MAY BECOME LOOSE AND INACCESSIBLE FOR POST-OPERATIVE ADJUSTMENTS, THUS REQUIRING REVISIONAL SURGERY."

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED, RAPIDPORT EZ ACCESS PORT FLIPPED 180 DEGREES." "COULD NOT ACCESS RAPIDPORT EZ ACCESS PORT DURING ROUTINE CLINIC VISIT, PATIENT SENT TO X-RAY AND PORT FLIP IDENTIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
312832 LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM WITH RAPIDPORT LTI ALLERGAN NA NI

Patients

Seq Age Sex Outcome Treatment
1 27 YR