INSTINCT ENDOSCOPIC HEMOCLIP
Report
- Report Number
- 1037905-2013-00507
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- June 4, 2013
- Report Date
- June 5, 2013
- Manufacturer
- WILSON-COOK MEDICAL INC.
- Product Code
- MND
- PMA / PMN Number
- K121505
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION EVALUATION: THE DEVICE WAS RETURNED WITHOUT THE CLIP THEREFORE A FUNCTION TEST COULD NOT BE PERFORMED. THE DRIVE WIRE IS SECURELY ATTACHED AT THE HANDLE AND THE HOOK OF THE DRIVE WIRE IS INTACT. A PRODUCT-SPECIFIC DISCREPANCY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THIS OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS. A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE CONDUCTED BECAUSE THE LOT NUMBER WAS NOT PROVIDED. INVESTIGATION CONCLUSIONS: RESISTANCE ENCOUNTERED WHILE ATTEMPTING TO DEPLOY THE CLIP AFTER THE CLIP IS CLOSED ON THE TISSUE (I.E. DIFFICULTY WITH CLIP RELEASE) CAN LEAD TO DAMAGE OF DEVICE COMPONENTS SUCH AS THE HOOK AT THE DISTAL END OF THE DRIVE WIRE OF THE SECURITY OF THE DRIVE WIRE TO THE HANDLE. THE INSTRUCTIONS FOR USE INSTRUCTS THE USER TO VERIFY SMOOTH HANDLE OPERATION AND CLIP OPERATION PRIOR TO USE. INSTRUCTIONS FOR USE STATES TO PERMANENTLY DEPLOY CLIP, PULL HANDLE SPOOL TOWARD HANDLE THUMB RING UNTIL CLIP DETACHES. NOTE: IF SEPARATION OF CLIP IS NOT IMMEDIATE, GENTLY MOVE CATHETER BACK AND FORTH OR USE OTHER ENDOSCOPIC MANEUVERS TO SEPARATE CATHETER FROM CLIP. INSTRUCTIONS FOR USE STATES IF CLIP DEPLOYMENT DEVICE IS USED WITH ENDOSCOPE IN A TORQUED OR RETROFLEXED POSITION, CLIP DEPLOYMENT DIFFICULTIES CAN OCCUR. PRIOR TO DISTRIBUTION, ALL INSTINCT ENDOSCOPIC HEMOCLIPS ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.
DURING AN UPPER ENDOSCOPY PROCEDURE FOR AN AVM (ARTERIOVENOUS MALFORMATION), THE PHYSICIAN SELECTED A COOK INSTINCT ENDOSCOPIC HEMOCLIP. THE CLIP WAS ATTACHED TO THE TISSUE SITE AND THE CLIP WOULD NOT RELEASE [FROM THE CLIP DEPLOYMENT DEVICE]. THE CLIP WAS UNABLE TO BE REOPENED FOR REMOVAL. THE PHYSICIAN PULLED THE CLIP OFF AND USED ANOTHER CLIP TO FINISH THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 244677 | INSTINCT ENDOSCOPIC HEMOCLIP | MND | WILSON-COOK MEDICAL INC. | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | OLYMPUS ENDOSCOPE 180 SERIES: MODEL UNK |