FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3221673 · Received July 1, 2013

Report

Report Number
1314492-2013-00955
Event Type
Malfunction
Date Received
July 1, 2013
Report Date
June 3, 2013
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE IN QUESTION. THE EVAL CONFIRMED A SYSTEM ERROR 322 CAUSED BY A FAILED UPPER LATCH SWITCH. THE FAILED UPPER LATCH SWITCH WAS REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP ALARMED FOR A SYSTEM ERROR 322 DURING AN INFUSION (MEDICATION, PROGRAMMED AMOUNT AND DELIVERY RATE UNK). IT IS UNK WHICH CARE AREA THE DEVICE WAS IN AT THE TIME OF THE ERROR, AND THERE WAS NO PT INJURY.