FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3221673
·
Received July 1, 2013
Report
- Report Number
- 1314492-2013-00955
- Event Type
- Malfunction
- Date Received
- July 1, 2013
- Report Date
- June 3, 2013
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE IN QUESTION. THE EVAL CONFIRMED A SYSTEM ERROR 322 CAUSED BY A FAILED UPPER LATCH SWITCH. THE FAILED UPPER LATCH SWITCH WAS REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PUMP ALARMED FOR A SYSTEM ERROR 322 DURING AN INFUSION (MEDICATION, PROGRAMMED AMOUNT AND DELIVERY RATE UNK). IT IS UNK WHICH CARE AREA THE DEVICE WAS IN AT THE TIME OF THE ERROR, AND THERE WAS NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 298115 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |