FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE AUTOMATIC MOIST HEAT PACK
MDR report key: 3221629
·
Received July 1, 2013
Report
- Report Number
- 1811605-2013-00092
- Event Type
- Malfunction
- Date Received
- July 1, 2013
- Date of Event
- June 13, 2013
- Report Date
- June 17, 2013
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE USER REPORTED THAT THE PAD SPARKED AND STOPPED WORKING. OUR INSPECTION FOUND A SMALL CUT IN THE CORD NEAR THE SWITCH THAT COULD HAVE EXPOSED THE USER TO BARE WIRE. CORD BREAKAGE USUALLY OCCURS WHEN THE CORD IS REPEATEDLY OVER BENT NEAR THE SWITCH. WE HAVE INITIATED A CAPA PROJECT ((B)(4)) TO REDUCE THE OCCURRENCE OF THIS ISSUE.
Description of Event or Problem · 1
THE USER REPORTED THAT THE PAD SPARKED THEN STOPPED WORKING.