FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3221495
·
Received July 3, 2013
Report
- Report Number
- 1314492-2013-00979
- Event Type
- Malfunction
- Date Received
- July 3, 2013
- Date of Event
- June 1, 2013
- Report Date
- June 4, 2013
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. A SYSTEM ERROR 322 WAS CONFIRMED THROUGH REVIEW OF THE DEVICE'S HISTORY LOG, AS WELL AS REPRODUCED DURING EVAL. THIS ERROR WAS CAUSED BY A FAILED LOWER LINK SWITCH. THE LINK SWITCH WAS FOUND TO HAVE CONTAMINATION ON THE SWITCH SPRING, THE METAL CONTACT AT THE BOTTOM OF THE SWITCH, AND ON THE BACKSIDE BETWEEN THE SWITCH AND THE FLEX. THE LOWER AUX WAS REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP WAS ALARMING CONSTANTLY FOR SYSTEM ERROR 322. IT WAS ALSO REPORTED THAT THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 303494 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|