FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3221474
·
Received July 3, 2013
Report
- Report Number
- 1314492-2013-00992
- Event Type
- Malfunction
- Date Received
- July 3, 2013
- Date of Event
- June 1, 2013
- Report Date
- June 5, 2013
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE IN QUESTION. TWO SYSTEM ERROR 322 EVENTS WERE CONFIRMED THROUGH REVIEW OF THE DEVICE HISTORY LOG. THE ALARM WAS UNABLE TO BE REPRODUCED DURING THE EVAL. THE DEVICE WAS TESTED FOR (B)(4) HOURS WITHOUT ANY OCCURRENCE OF SYSTEM ERROR 322. THE UPPER AND LOWER AUX ARE KNOWN CONTRIBUTE TO THIS ALARM AND HAVE BEEN REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP WAS HAVING SYSTEM ERROR 322 ALARMS. IT WAS ALSO REPORTED THAT THERE WAS NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 303440 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |