LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-01755
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- May 27, 2013
- Report Date
- June 28, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 102) WAS CONFIRMED. UPON INVESTIGATION RESISTOR R45 WAS OPEN ON THE C/A BOARD, WHICH CAUSED THE SERVICE CODE 102. THE CAUSE FOR THE OPEN R45 RESISTOR WAS A BROKEN LEAD ON HIGH-VOLTAGE CAPACITOR C22 ON THE DEFIBRILLATOR BOARD. THE ROOT CAUSE FOR THE BROKEN LEAD ON C22 COULD NOT BE POSITIVELY IDENTIFIED, BUT THE DAMAGE LIKELY RESULTED FROM THE MONITOR IMPACTING A HARD SURFACE. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. A MECHANICAL DESIGN CHANGE TO REDUCE THE PCA STRAIN WAS APPROVED BY FDA ON (B)(4) 2012. IMPLEMENTATION BEGAN ON (B)(4) 2013. RESULTS WILL BE MONITORED. NO ADVERSE EVENT RESULTED FROM THE BROKEN LEAD ON C22. THE PATIENT RECEIVED A REPLACEMENT MONITOR.
A (B)(6) MALE PATIENT'S WIFE CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A SERVICE CODE 102 (CHARGE PROFILE FAULT). THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 301492 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |