HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2013-18052
- Event Type
- Malfunction
- Date Received
- July 11, 2013
- Date of Event
- June 20, 2013
- Report Date
- June 20, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NURSE
Narratives
(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). A REVIEW OF ALL BATCH RECORD DOCUMENTS FOR POTENTIALLY ASSOCIATED LOT NUMBER H12L15042 WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SUBMITTED.
IT WAS REPORTED THAT DURING TROUBLESHOOTING FOR THE UNRELATED ALARM THE REGISTERED NURSE (RN) NOTICED AIR IN THE PATIENT LINE DURING THE INITIAL DRAIN, WHILE THE HOME PATIENT (HP) WAS CONNECTED. DURING TROUBLESHOOTING NOTHING UNUSUAL WAS FOUND THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) HAD THE RN CLOSE THE CLAMPS, DISCONNECT THE HP AND CYCLE THE POWER. THE TSR ASSISTED THE RN WITH ENDING THE THERAPY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 319929 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME | H12L15042 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | HOMECHOICE |