FDA Adverse Event Malfunction Summary report: N

STERLING?

MDR report key: 3221390 · Received July 11, 2013

Report

Report Number
2134265-2013-04764
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
June 14, 2013
Report Date
June 14, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
DQY
PMA / PMN Number
K053116
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR.: THE COMPLAINT DEVICE WAS RECEIVED INSIDE A CARRIER TUBE (HOOP) WITHIN A STERLING SHELF BOX THAT MATCHED THE INFORMATION ON THE DEVICE. MAGNIFIED INSPECTION REVEALED NO DAMAGE TO THE CATHETER. THE DEVICE WAS INFLATED TO RATED BURST PRESSURE (RBP) WITH AN INFLATION DEVICE FILLED WITH WATER. THE DEVICE MAINTAINED RBP FOR FIVE (5) MINUTES WITH NO INDICATION OF ANY LEAKS OR OTHER IRREGULARITIES. FUNCTIONAL TESTING REVEALED NO EVIDENCE OF THE ALLEGED BALLOON BURST. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS NOT CONFIRMED. (B)(4).

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN THE SEVERELY TORTUOUS AND MODERATELY CALCIFIED FOREARM ARTERY. A 6.0MM X 40MM X 80CM STERLING¿ BALLOON CATHETER WAS ADVANCED TO THE TARGET LESION. UPON THIRD INFLATION AT 12 ATMOSPHERES, THE BALLOON RUPTURED. PROCEDURE WAS COMPLETED WITH ANOTHER SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT STATUS IS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN THE SEVERELY TORTUOUS AND MODERATELY CALCIFIED FOREARM ARTERY. A 6.0MM X 40MM X 80CM STERLING BALLOON CATHETER WAS ADVANCED TO THE TARGET LESION. UPON THIRD INFLATION AT 12 ATMOSPHERES, THE BALLOON RUPTURED. PROCEDURE WAS COMPLETED WITH ANOTHER SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320661 STERLING? CATHETER, PERCUTANEOUS DQY BOSTON SCIENTIFIC - MAPLE GROVE H74939032604080 13837045

Patients

Seq Age Sex Outcome Treatment
1