ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-10333
- Event Type
- Injury
- Date Received
- July 11, 2013
- Report Date
- June 12, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
DEVICE EVALUATION FOLLOW-UP #1 SUBMITTED 09/11/2013: THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON 08/25/2013 WITH THE FOLLOWING RESULTS: TESTING WAS UNABLE TO DUPLICATE THE COMPLAINT. NO DEFECT WAS FOUND. A 29HR FLOW ACCURACY TEST WAS PERFORMED; PUMP PASSED AND WAS FOUND TO BE DELIVERING WITHIN THE REQUIRED SPECIFICATIONS. NO ALARMS RELATED TO THE COMPLAINT NOTED IN THE ALARM HISTORY; ONLY TYPICAL USAGE OBSERVED. THE BASAL AND BOLUSES ADD UP APPROPRIATELY TO TOTAL THE TDD; SHOWING THE PUMP WAS DELIVERING ACCURATELY UNTIL THE LAST DATE USED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER, THE PATIENT'S MOTHER, CONTACTED ANIMAS AND ALLEGED THAT THE PATIENT'S BLOOD GLUCOSE (BG) IS 561 MG/DL, WITH MILD NAUSEOUS. MOM STATES SHE BELIEVES THAT PATIENT GAVE ALL BOLUSES AS INTENDED. PATIENT TOOK A PHENERGAN FOR THE NAUSEA AND MOM GAVE A CORRECTION BOLUS VIA PUMP. MOM REPORTS PUMP HISTORY DOES NOT RECORD ABOUT EVERY BOLUS. CUSTOMER SUPPORT (CS) ADVISED MOM TO DISCONTINUE PUMP UNTIL REPLACEMENT PUMP ARRIVES, TO USE AN ALTERNATE FORM OF INSULIN DELIVERY, AND TO CALL BACK IF BG DOES NOT COME DOWN. THIS COMPLAINT IS BEING REPORTED AS A PATIENT ON PUMP THERAPY EXPERIENCED HYPERGLYCEMIA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320141 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR | Life Threatening |