FDA Adverse Event Malfunction Summary report: N

CERVIC-DISTRACT R W/ADJUST-ANGLE

MDR report key: 3221040 · Received July 11, 2013

Report

Report Number
8030965-2013-04282
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
May 8, 2013
Report Date
June 14, 2013
Manufacturer
SYNTHES GMBH
Product Code
HST
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. THE INVESTIGATION OF THE COMPLAINED CERVICAL DISTRACTOR CONFIRMED THAT ONE ARM IS BROKEN OFF AT THE JOINT FOR THE VARIABLE ANGLE. THE LCP DRILL BIT WAS ANALYZED FOR CONFORMANCE TO PRINT SPECIFICATIONS AS WELL AS THE DEVICE HISTORY RECORD RESEARCHED AND NO ABNORMAL FINDINGS WERE IDENTIFIED. THE BROKEN SURFACE IS HOMOGENOUS WHICH INDICATES MATERIAL CONFORMITY AS WELL. THE CAUSE IS UNKNOWN. THIS IS INDICATIVE OF TOO MUCH MECHANICAL FORCE APPLIED DURING THE SURGERY. NO PRODUCT FAULT COULD BE DETECTED.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: THE INSTRUMENT BROKE IN THE BEGINNING OF AN ANTERIOR CERVICAL DECOMPRESSION AND FUSION (ACDF) SURGERY ON (B)(6) 2013 (CERVIOS CAGE). THERE WAS NO HARM FOR THE PATIENT OR THE OPERATOR. THE INCIDENT DELAYED THE OPERATION FOR ABOUT 5 MINUTES. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319505 CERVIC-DISTRACT R W/ADJUST-ANGLE HST SYNTHES GMBH T953280

Patients

Seq Age Sex Outcome Treatment
1