FLEXCATH STEERABLE SHEATH

MDR report key: 3221016 · Received July 11, 2013

Report

Report Number
3002648230-2013-00107
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
October 17, 2012
Report Date
April 23, 2013
Manufacturer
MEDTRONIC CRYOCATH LP
Product Code
DRA
PMA / PMN Number
K081049
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTION OF THE RETURNED DEVICE SHOWED THAT THE SHAFT WAS INTACT WITH NO APPARENT ISSUES. THE REPORTED AIR INGRESS COULD NOT BE REPRODUCED. MANY ASPIRATION / INJECTIONS WERE PERFORMED WITHOUT HAVING AIR BUBBLES OR LEAKS; THE HEMOSTATIC VALVE WAS LEAK TIGHT. THE SHEATH PASSED THE INSPECTION AS PER SPECIFICATION. THIS REPORT WILL BE RECORDED AND TRENDED. A FIELD ACTION WAS INITIATED IN (B)(4) 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THE POTENTIAL LEAK IN THE HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS, WHICH CAN RESULT IN AIR INGRESS DURING ASPIRATION. THE INITIAL DECISION FOR THIS EVENT WAS NOT MDR REPORTABLE SINCE THE SHEATH PASSED INSPECTION AS PER SPECIFICATION. DECISION CHANGED TO REPORTABLE FOLLOWING REVISION TO REGULATORY REPORTING CRITERIA EFFECTIVE (B)(4) 2013.

Description of Event or Problem · 1

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THERE WAS AIR INGRESS DURING ASPIRATION OF THE SHEATH. THE SHEATH WAS REPLACED AND NO FURTHER ISSUE WAS NOTED. NO ADVERSE EVENT OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319476 FLEXCATH STEERABLE SHEATH CATHETER, STEERABLE DRA MEDTRONIC CRYOCATH LP 3FC12

Patients

Seq Age Sex Outcome Treatment
1