PWRD ECH FLEX 60MM
Report
- Report Number
- 3005075853-2013-03515
- Event Type
- Malfunction
- Date Received
- July 11, 2013
- Date of Event
- June 13, 2013
- Report Date
- June 18, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K110385
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH AN ECR60G RELOAD LOADED IN THE DEVICE. THE CARTRIDGE RELOAD WAS RECEIVED PARTIALLY FIRED 1/10. IT IS POSSIBLE THAT WHILE LOADING THE RELOAD, THE CARTRIDGE WAS PUSHED FARTHER BACK THAN THE CARTRIDGE ALIGNMENT STOP WINDOWS RESULTING IN THE KNIFE PUSHING THE ONE PIECE SLED FORWARD AND LOCKING THE CARTRIDGE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND IT FIRED, AND FORMED ALL THE STAPLES AS INTENDED; THE CUT AND STAPLE LINE WERE COMPLETE. HOWEVER, THE KNIFE WAS NOTED NICKED. ONE POSSIBLE CAUSE FOR THIS TYPE OF DAMAGE TO THE KNIFE IS WHEN THE DEVICE IS FIRED OVER AN ALREADY EXISTING STAPLE LINE, HARD OBJECT OR THICKER TISSUE THAN INDICATED. REPEATEDLY FIRING ACROSS EXISTING STAPLE LINES CAN ALSO REDUCE THE ABILITY TO CUT CLEANLY. TO MITIGATE THE POTENTIAL FOR STAPLES GETTING INTO THE CARTRIDGE AND INTERFERING WITH THE KNIFE PATH DURING DEVICE FIRING, PRIOR TO RELOADING THE DEVICE, RINSE THE ANVIL AND CARTRIDGE JAW IN STERILE SOLUTION AND THEN WIPE THE ANVIL AND CARTRIDGE JAW TO CLEAN ANY FORMED BUT UNUSED STAPLES FROM THE DEVICE. ADDITIONALLY, PROPER CARE SHOULD BE TAKEN WHEN PLACING THE DEVICE ON THE TISSUE TO BE STAPLED, TO ENSURE THAT NO HARD OBSTRUCTION SUCH AS A CLIP IS INCLUDED WITH THE TISSUE INSIDE THE JAWS. PLEASE NOTE IF THE INSTRUMENT IS PARTIALLY FIRED, SLIDE THE KNIFE REVERSE SWITCH FORWARD TO RETURN THE KNIFE TO HOME POSITION. TO OPEN THE JAWS, SQUEEZE THE CLOSING TRIGGER, AND THEN SIMULTANEOUSLY PRESS THE ANVIL RELEASE SWITCH ON EITHER SIDE OF THE INSTRUMENT. WHILE PRESSURE IS STILL ON THE ANVIL RELEASE SWITCH, SLOWLY RELEASE THE CLOSING TRIGGER. AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING A SLEEVE GASTRECTOMY, THE STAPLE FROZE ON THE FIFTH FIRING THE DEVICE WAS UNABLE TO BE OPENED; COULD NOT OVERRIDE. THE BATTERY WAS TAKEN OUT AND PUT BACK IN AND THE REVERSE BUTTON WOULD NOT WORK. THE DEVICE WAS NOT IN THE PATIENT OR ON THE TISSUE. A SECOND DEVICE WAS OPENED AND THE CASE WAS COMPLETED WITH NO PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320342 | PWRD ECH FLEX 60MM | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | K4CM24 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |