FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3220920 · Received July 11, 2013

Report

Report Number
1416980-2013-18049
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
May 16, 2013
Report Date
June 17, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING THE EVENT HISTORY LOG REVIEW, AN INCREASED INTRA-PERITONEAL VOLUME EVENT WAS IDENTIFIED. THE ROOT CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE FALSE EMPTY DETECT AND USE ERROR, CLINICIAN INAPPROPRIATELY SET MINIMUM DRAIN VOLUME PERCENT SETTING TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER'S GUIDE PROVIDES A WARNING "IF YOU SET THE MINIMUM DRAIN VOLUME % TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR THE PATIENT. THIS COULD RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ANY RELEVANT INFORMATION IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 20:56:34. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 2029ML, INDICATING THE HOME PATIENT (HP) DRAINED 2029ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321116 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 48 YR