FDA Adverse Event Malfunction Summary report: N

T5 LT WT FIBER OPT SURG HELMET

MDR report key: 3220911 · Received July 11, 2013

Report

Report Number
0001811755-2013-01623
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
June 18, 2013
Report Date
June 18, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
FXZ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED. FAILURE ANALYSIS IS IN PROGRESS.

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, IT WAS DETERMINED THAT HEAT BUILD-UP FROM A DAMAGED FIBER OPTIC CABLE WAS A PROBABLE CAUSE OF THE REPORTED EVENT. THE DEVICE WAS NOT RETURNED TO THE USER FACILITY, AS IT IS NOT A REPAIRABLE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY, THE FIBER OPTIC CABLE PLUG WAS FOUND TO BE MELTED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY, THE FIBER OPTIC CABLE PLUG WAS FOUND TO BE MELTED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.