- Report Number
- 1416980-2013-18046
- Event Type
- Malfunction
- Date Received
- July 11, 2013
- Date of Event
- June 17, 2013
- Report Date
- June 17, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
(B)(4). A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). EVALUATION SUMMARY: THE SAMPLE WAS RETURNED FOR EVALUATION. A VISUAL TEST IDENTIFIED THE SEPARATED TUBING; A SOLVENT RING WAS OBSERVED AT THE END OF TUBING. A CAUSE FOR THE REPORTED CONDITION COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
IT WAS REPORTED THAT THE TUBING SEPARATED FROM THE MALE LEUR OF A CATHETER EXTENSION SET JUST AFTER THE START OF INFUSION. THERE WAS PATIENT INVOLVEMENT; HOWEVER, THERE WAS NOT PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.