CD HORIZON SPINAL SYSTEM
Report
- Report Number
- 1030489-2013-03003
- Event Type
- Malfunction
- Date Received
- July 11, 2013
- Date of Event
- June 12, 2013
- Report Date
- June 12, 2013
- Manufacturer
- WARSAW ORTHOPEDICS
- Product Code
- KWP
- PMA / PMN Number
- K050439
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(6). (B)(4). VISUAL AND OPTICAL INSPECTION OF THE CMAS BONE SCREW HEX NOTED SOME MATERIAL DEFORMATION OF THE HEX. FUNCTIONAL EVALUATION OF THE CMAS SCREW WITH A SAMPLE CMAS DRIVER AND SIMULATED BONE DID NOT IDENTIFY ANY ISSUE WITH FULLY DRIVING THE SCREW INTO THE SIMULATED BONE. UNABLE TO REPLICATE CUSTOMER CONCERN; AFTER VISUAL, OPTICAL AND FUNCTIONAL EVALUATION, NO EVIDENCE WAS FOUND THAT WOULD SUGGEST A DEFECT IN MANUFACTURING OR PROCESSING OF THE IMPLANT OR ASSOCIATED COMPONENTS. THE IMPLANT APPEARS TO BE CAPABLE OF PERFORMING ITS INTENDED FUNCTION.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A POSTERIOR SPINAL SURGICAL PROCEDURE. IT WAS REPORTED THAT THE ILIAC SCREW STRIPPED DURING INSERTION. THE SCREW WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.