OT ULTRALINK METER
Report
- Report Number
- 3008382007-2013-19332
- Event Type
- Malfunction
- Date Received
- July 11, 2013
- Report Date
- June 29, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K073231
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 (08/12/2013)-PRODUCT EVALUATION: THE SUBJECT METER WAS RETURNED ON (B)(4) 2013 AND ANALYSIS COMPLETED ON (B)(4) 2013 BY LIFESCAN PRODUCT ANALYSIS. THE ANALYSIS OF THE METER FOUND THAT THE CRYSTAL X1 WAS DEFECTIVE. HIGH STANDBY CURRENT WAS OBSERVED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING DOES NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.