FDA Adverse Event Injury Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 3220408 · Received July 11, 2013

Report

Report Number
2938836-2013-04261
Event Type
Injury
Date Received
July 11, 2013
Date of Event
May 29, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
P950022
Removal / Correction Number
Z0458
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED FIELD EVENT OF HIGH LEAD IMPEDANCE WAS NOT CONFIRMED. EXTERNAL INSULATION ABRASION WAS NOTED AT 14.6- 16.0CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO A FEATURE OF THE HEART. THE ETFE COATING WAS ABRADED AT THE SAME LOCATION. EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE NOTED AT 12.0-14.6CM FROM THE DISTAL TIP. THE ETFE COATING WAS ABRADED AT THE SAME LOCATION.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS EXPLANTED DUE TO INCREASED PACING LEAD IMPEDANCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321074 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7001/65 NA

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention