FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 3220407 · Received July 11, 2013

Report

Report Number
2938836-2013-04181
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
May 14, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4). (B)(60. FAILURE (EVENT) OBSERVED DURING ANALYSIS. INTERNAL INSULATION ABRASIONS NOTED AT 15.0-15.02CM, 15.2-15.22CM, 42.2-42.5CM, AND 42.3- 42.35CM FROM THE ELECTRODE TIP. THE ETFE COATING WAS INTACT AT ALL ABRASION LOCATIONS.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321017 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1581/60 NA

Patients

Seq Age Sex Outcome Treatment
1