FDA Adverse Event Malfunction Summary report: N

ELLIPSE DR

MDR report key: 3220365 · Received July 11, 2013

Report

Report Number
2938836-2013-04494
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
May 8, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED IN CLINIC FOR A FOLLOW UP, POST-PACED T-WAVE OVERSENSING WAS OBSERVED. THE PATIENT DID NOT RECEIVE THERAPY. THE OVERSENSING WAS NOTED ON A STORED EGM. THE DEVICE WAS REPROGRAMMED SUCCESSFULLY. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320954 ELLIPSE DR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD2311-36 NA

Patients

Seq Age Sex Outcome Treatment
1 53 YR