FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA CRT-D

MDR report key: 3220363 · Received July 11, 2013

Report

Report Number
2938836-2013-04653
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
May 29, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT ONE DAY POST IMPLANT, INTERROGATION REVEALED AN ALERT FOR NONSUSTAINED LEAD NOISE DUE TO UNDERSENSING OF PVCS ON THE DISCRIMINATION CHANNEL. REVIEW OF EGMS SHOWED NO NOISE. DEVICE WAS REPROGRAMMED AND WILL BE MONITORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320503 QUADRA ASSURA CRT-D IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3267-40 NA

Patients

Seq Age Sex Outcome Treatment
1 66 YR