FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM PASSIVE FIXATION, DF4 CONNECTOR

MDR report key: 3220333 · Received July 11, 2013

Report

Report Number
2938836-2013-04398
Event Type
Injury
Date Received
July 11, 2013
Date of Event
May 29, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EXTERNAL INSULATION ABRASION WAS NOTED AT 9.0-9.5CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS ABRADED AT THE SAME LOCATION. THIS IS CONSISTENT WITH THE OBSERVATION FROM THE FIELD OF NOISE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING INTERROGATION NOISE WAS OBSERVED ON RV LEAD. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320338 DURATA STS OPTIM PASSIVE FIXATION, DF4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7170Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention