DURATA STS OPTIM PASSIVE FIXATION, DF4 CONNECTOR
MDR report key: 3220333
·
Received July 11, 2013
Report
- Report Number
- 2938836-2013-04398
- Event Type
- Injury
- Date Received
- July 11, 2013
- Date of Event
- May 29, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EXTERNAL INSULATION ABRASION WAS NOTED AT 9.0-9.5CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS ABRADED AT THE SAME LOCATION. THIS IS CONSISTENT WITH THE OBSERVATION FROM THE FIELD OF NOISE.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING INTERROGATION NOISE WAS OBSERVED ON RV LEAD. THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320338 | DURATA STS OPTIM PASSIVE FIXATION, DF4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7170Q/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |