FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 3220326 · Received July 11, 2013

Report

Report Number
2938836-2013-04322
Event Type
Injury
Date Received
July 11, 2013
Date of Event
May 7, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE LEAD WAS OTHERWISE NORMAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT HIGH, OUT OF RANGE HV LEAD IMPEDANCE WAS OBSERVED. DFT TESTING WAS PERFORMED AND COULD NOT BE TERMINATED BY THE ICD. EXTERNAL DEFIBRILLATION WAS USED. THE LEAD AND DEVICE WERE EXPLANTED. THE PATIENT CONDITION IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319624 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7121/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention (B)(4), 562434