FDA Adverse Event
Malfunction
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 3220268
·
Received July 11, 2013
Report
- Report Number
- 2938836-2013-04193
- Event Type
- Malfunction
- Date Received
- July 11, 2013
- Date of Event
- May 15, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4). (B)(6). FAILURE (EVENT) OBSERVED DURING ANALYSIS. EXTERNAL INSULATION ABRASION WAS FOUND AT 6.6-8.2CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE. INTERNAL INSULATION ABRASION WAS FOUND AT 11.4-13.8CM FROM THE DISTAL TIP. INTERNAL INSULATION ABRASION UNDER THE SVC SHOCK COIL WAS FOUND AT 27.6-28.0CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321105 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1581/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |