FDA Adverse Event
Injury
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 3220265
·
Received July 11, 2013
Report
- Report Number
- 2938836-2013-04182
- Event Type
- Injury
- Date Received
- July 11, 2013
- Date of Event
- May 29, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0457
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EXTERNAL INSULATION ABRASION WAS NOTED AT 0.4-1.1CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THE SAME LOCATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED TO THE OR FOR LEAD REPLACEMENT DUE TO NOISE. EXTERNALIZED CONDUCTORS WERE NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321104 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1581/60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention |