RIATA ACTIVE FIXATION

MDR report key: 3220265 · Received July 11, 2013

Report

Report Number
2938836-2013-04182
Event Type
Injury
Date Received
July 11, 2013
Date of Event
May 29, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EXTERNAL INSULATION ABRASION WAS NOTED AT 0.4-1.1CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THE SAME LOCATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED TO THE OR FOR LEAD REPLACEMENT DUE TO NOISE. EXTERNALIZED CONDUCTORS WERE NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321104 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1581/60 NA

Patients

Seq Age Sex Outcome Treatment
1 70 YR Required Intervention