FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 3220261 · Received July 11, 2013

Report

Report Number
2938836-2013-04171
Event Type
Injury
Date Received
July 11, 2013
Date of Event
May 6, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(6). EXTERNAL INSULATION ABRASION WAS NOTED AT 7.4-8.4CM FROM THE END OF THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS ABRADED AT THE SAME LOCATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED DUE TO SUSPECTED CLAVICLE CRUSH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319433 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1580/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention