FDA Adverse Event Malfunction Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 3220255 · Received July 11, 2013

Report

Report Number
2938836-2013-04350
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
March 25, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CLINIC RECEIVED HIGH IMPEDANCE MEASUREMENT TRANSMISSION VIA A COMPETITORS SYSTEM. THERE WAS NO INAPPROPRIATE THERAPY. THE LEAD WILL CONTINUE TO BE MONITORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319431 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7122/65 NA

Patients

Seq Age Sex Outcome Treatment
1 38 YR