FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR

MDR report key: 3220218 · Received July 11, 2013

Report

Report Number
2938836-2013-04362
Event Type
Injury
Date Received
July 11, 2013
Date of Event
April 16, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED SEVERAL SHOCKS. REVIEW OF EGMS SHOWED OVERSENSING. X-RAY SHOWED A LEAD DISLODGEMENT. THE LEAD WAS EXPLANTED AND REPLACED WHEN ATTEMPTED REPOSITIONING WAS UNSUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321006 DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7122Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention