DURATA STS OPTIM ACTIVE FIXATION
Report
- Report Number
- 2938836-2013-04342
- Event Type
- Injury
- Date Received
- July 11, 2013
- Date of Event
- April 22, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
EXTERNAL INSULATION ABRASION WAS NOTED AT 38.0-38.3CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE SILICONE INSULATION WAS INTACT AT THE SAME LOCATION.
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED EXTERNAL INSULATION ABRASION WAS NOTED AT 38.0-38.3CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THIS LOCATION.
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC AFTER EXPERIENCING POCKET STIMULATION. RV THRESHOLDS WERE INCREASED, R-WAVES HAD DECREASED AND IMPEDANCE WAS TRENDING DOWN. NOISE WAS REPRODUCIBLE DURING ARM MOVEMENTS. THE POCKET WAS OPENED AND AN INSULATION ABRASION ON THE PACE/SENSE PORTION OF THE LEAD WAS OBSERVED. DURING LASER LEAD EXPLANT PROCEDURE, THE ATRIAL LEAD PULLED BACK AND WAS REPLACED AS WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321298 | DURATA STS OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7122/60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR | Required Intervention |