FDA Adverse Event Malfunction Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 3220178 · Received July 11, 2013

Report

Report Number
2938836-2013-04248
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
May 24, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). FAILURE (EVENT) OBSERVED DURING ANALYSIS. A PARTIAL LEAD WITH 46.2CM OF THE DISTAL END WAS RETURNED. EXTERNAL INSULATION ABRASION WAS FOUND AT 40.2-41.1CM FROM THE HELIX, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THE SAME LOCATION.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320268 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7000/65 NA

Patients

Seq Age Sex Outcome Treatment
1