UNIFY QUADRA, CRT-D, DF-4 CONNECTOR
Report
- Report Number
- 2938836-2013-04550
- Event Type
- Injury
- Date Received
- July 11, 2013
- Date of Event
- May 21, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NIK
- PMA / PMN Number
- P030054
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
NO COMMUNICATION COULD BE ESTABLISHED BETWEEN THE DEVICE AND THE PROGRAMMER UPON RECEIPT. LOW BATTERY VOLTAGE WAS OBSERVED. WITH AN EXTERNAL POWER SUPPLY ATTACHED, THE DEVICE FUNCTIONED NORMALLY; NO HIGH CURRENT WAS DETECTED DURING TESTING AND THE POWER CONSUMPTION WAS NORMAL. A LONGEVITY CALCULATION WAS PERFORMED AND WAS FOUND TO BE BELOW THE EXPECTED LIMITS. THE ORIGINAL BATTERY WAS SENT TO THE VENDOR FOR FURTHER EVALUATION AND NO ANOMALY WAS FOUND. THE CAUSE OF THE BATTERY DEPLETION REMAINS UNDETERMINED.
IT WAS REPORTED THAT NO CAPTURE WAS OBSERVED AND THE DEVICE COULD NOT BE INTERROGATED. THE DEVICE WAS EXPLANTED AND REPLACED.