FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA CRT-D, DF4 CONNECTOR

MDR report key: 3220140 · Received July 11, 2013

Report

Report Number
2938836-2013-04625
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
May 24, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT (B)(6) TRANSMISSION FOR NON-SUSTAINED LEAD NOISE WAS RECEIVED. REVIEW OF EGMS REVEALED THE NON-SUSTAINED LEAD NOISE WAS CAUSED BY AN INTERMITTENT ATRIAL UNDERSENSING. PROGRAMMING CHANGES WE RE RECOMMENDED. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321063 QUADRA ASSURA CRT-D, DF4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3265-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 60 YR