FDA Adverse Event Malfunction Summary report: N

FORTIFY ASSURA DR

MDR report key: 3220058 · Received July 11, 2013

Report

Report Number
2938836-2013-04463
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
April 24, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED VIA MERLIN.NET TRANSMISSION. NONSUSTAINED LEAD NOISE AND POST PACED T WAVE OVERSENSING WERE OBSERVED ON STORED EGMS. PROGRAMMING CHANGES WERE RECOMMENDED. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319538 FORTIFY ASSURA DR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD2257-40 NA

Patients

Seq Age Sex Outcome Treatment
1 82 YR