FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA CRT-D, DF4 CONNECTOR

MDR report key: 3220047 · Received July 11, 2013

Report

Report Number
2938836-2013-04645
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
May 2, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT NON-SUSTAINED LEAD NOISE EPISODES WERE STORED FOR POST-PACED T-WAVE OVERSENSING VIA A MERLIN.NET TRANSMISSION. FAR-R-WAVE OVERSENSING ON THE ATRIAL CHANNEL WAS ALSO NOTED. PROGRAMMING CHANGES WERE RECOMMENDED. THE DEVICE REMAINS IMPLANTED AND PATIENT WILL BE MONITORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321312 QUADRA ASSURA CRT-D, DF4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3265-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 51 YR (B)(4)