FDA Adverse Event
Malfunction
Summary report: N
QUADRA ASSURA CRT-D, DF4 CONNECTOR
MDR report key: 3220045
·
Received July 11, 2013
Report
- Report Number
- 2938836-2013-04640
- Event Type
- Malfunction
- Date Received
- July 11, 2013
- Date of Event
- April 26, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NIK
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED TO THE CLINIC FOR A FOLLOW-UP. THE DEVICE WAS POST-PACED T-WAVE OVERSENSING. THE OVERSENSING WAS NOTED ON A REALTIME EGM. PROGRAMMING CHANGES WERE RECOMMENDED. THE DEVICE REMAINS IMPLANTED.
Description of Event or Problem · 1
NEW INFORMATION NOTES THE DEVICE WAS SUCCESSFULLY REPROGRAMMED. THE PATIENT HAS HAD NO FURTHER PROBLEMS FOR T-WAVE OVERSENSING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320292 | QUADRA ASSURA CRT-D, DF4 CONNECTOR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | CD3265-40Q | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |