FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA CRT-D, DF4 CONNECTOR

MDR report key: 3220028 · Received July 11, 2013

Report

Report Number
2938836-2013-04638
Event Type
Malfunction
Date Received
July 11, 2013
Date of Event
April 5, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THE REPORTED FIELD EVENT OF NOISE WAS CONFIRMED TO OCCUR AFTER A CAPACITOR MAINTENANCE CHARGE. THE CAUSE OF NOISE WAS CURRENT LEAKAGE THROUGH THE HV OUTPUT SWITCHES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS OVERSENSING NOISE WHILE IN THE BOX BEFORE AN IMPLANT PROCEDURE. THE DEVICE WAS NOT IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319813 QUADRA ASSURA CRT-D, DF4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3265-40Q NA

Patients

Seq Age Sex Outcome Treatment
1