DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR
MDR report key: 3220013
·
Received July 11, 2013
Report
- Report Number
- 2938836-2013-04306
- Event Type
- Injury
- Date Received
- July 11, 2013
- Date of Event
- April 3, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN ATTEMPTED LEAD REVISION, THE SCREW OF THE LEAD WOULD NOT LOOSEN. THE PHYSICIAN TURNED THE SCREW MULTIPLE TIMES. THE SCREW RETRACTED INTO THE LUMEN OF THE LEAD. THE MANDRIN WAS STUCK IN THE LEAD. HIGH DEFIBRILLATION THRESHOLDS WERE NOTED. THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321090 | DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7120Q/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |