FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR

MDR report key: 3220013 · Received July 11, 2013

Report

Report Number
2938836-2013-04306
Event Type
Injury
Date Received
July 11, 2013
Date of Event
April 3, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ATTEMPTED LEAD REVISION, THE SCREW OF THE LEAD WOULD NOT LOOSEN. THE PHYSICIAN TURNED THE SCREW MULTIPLE TIMES. THE SCREW RETRACTED INTO THE LUMEN OF THE LEAD. THE MANDRIN WAS STUCK IN THE LEAD. HIGH DEFIBRILLATION THRESHOLDS WERE NOTED. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321090 DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120Q/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention