FDA Adverse Event Injury Summary report: N

RIATA ST OPTIM ACTIVE FIXATION

MDR report key: 3220009 · Received July 11, 2013

Report

Report Number
2938836-2013-04268
Event Type
Injury
Date Received
July 11, 2013
Date of Event
April 26, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

A PARTIAL LEAD MEASURING 52.4CM FROM THE DISTAL TIP WAS RETURNED FOR EVALUATION. INTERNAL INSULATION ABRASION WAS NOTED AT 29.6-29.9CM FROM THE DISTAL TIP. INTERNAL INSULATION ABRASION UNDER THE RV SHOCK COIL WAS NOTED AT 5.0 -6.5CM FROM THE DISTAL TIP. INTERNAL INSULATION ABRASION UNDER THE RV SHOCK COIL WAS NOTED AT 29.2-29.3CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. INTERNAL INSULATION ABRASION WAS NOTED UNDER THE SVC SHOCK COIL AT 24.6-26. 6CM FROM THE DISTAL TIP. ONE RV CONDUCTOR AND THE SVC SHOCK COIL WERE MELTED AT 24.8CM FROM THE DISTAL TIP. THIS IS CONSISTENT WITH THE FIELD COMPLAINT OF LOW HV LEAD IMPEDANCE.

Description of Event or Problem · 1

LOW HV LEAD IMPEDANCE WAS OBSERVED. IT WAS REPORTED THAT PRIOR TO EXPLANT, X-RAY DID NOT SHOW ANY ANOMALIES. UPON EXPLANT, IT WAS REPORTED THAT THE LEAD WAS CUT AND EXTERNALIZATION WAS OBSERVED. SINCE NO ANOMALIES WERE OBSERVED PRIOR TO EXPLANT, IT IS LIKELY THAT THE REPORTED EXTERNALIZATION OBSERVED UP EXPLANT WAS DUE TO THE LEAD HAVING BEEN CUT DURING EXPLANT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319776 RIATA ST OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7021/65 NA

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention