FDA Adverse Event Other Summary report: N

T-FLEX ASPHERIC

MDR report key: 3214097 · Received July 3, 2013

Report

Report Number
9611165-2013-00057
Event Type
Other
Date Received
July 3, 2013
Report Date
July 2, 2013
Manufacturer
RAYNER INTRAOCULAR LENSES LTD.
Product Code
HQL
PMA / PMN Number
P060011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RAYNER MADE INITIAL CONTACT WITH ALL CUSTOMERS WHERE AFFECTED BATCHES HAD BEEN SENT IN ORDER TO DETERMINE WHETHER THE IDENTIFIED MODELS HAD BEEN USED OR REMAINED IN THEIR STOCKS. FOLLOWING COMMUNICATION WITH ALL AFFECTED CUSTOMERS RAYNER RECEIVED CONFIRMATION THAT NO LENSES RELEASED TO MARKET HAD BEEN USED. ALL LENSES IN THE POSSESSION OF CUSTOMERS WERE QUARANTINED AND RETURNED TO RAYNER. RAYNER HAS ACHIEVED 100% DEVICE RECONCILIATION. DEVICES FROM ALL AFFECTED MARKETS WERE RECONCILED IN THEIR ENTIRETY BY 24TH JUNE 2013. THE POUCH MFR HAS ATTRIBUTED A LEAK IN THE SEALING MACHINE DURING PRODUCTION AS THE ROOT CAUSE OF THE SEAL WEAKNESS. THIS EVENT HAS OCCURRED AS A RESULT OF A MFG DEFICIENCY ON THE PART OF THE POUCH SUPPLIER. RAYNER IN COLLABORATION WITH THE POUCH MFR IS WORKING TO DETERMINE A CORRECTIVE ACTION PLAN. (B)(4).

Description of Event or Problem · 1

IN (B)(6) 2013, RAYNER INTRAOCULAR LENSES LIMITED IDENTIFIED A POTENTIAL WEAKNESS OF THE OUTER PAPER POUCH CHEVRON SEAL USED TO ENCASE THE PRIMARY PACK (THE IOL BLISTER POUCH. RAYNER DETERMINED THAT AS A RESULT OF THIS WEAKNESS THE STERILITY OF THE PRIMARY PACK MAY BE COMPROMISED. AN INVESTIGATION WAS PERFORMED BY RAYNER TO DETERMINE THE EXTENT OF THE ISSUE IN ORDER TO IDENTIFY THE SPECIFIC POUCH RANGE AFFECTED. THIS INCLUDED A REVIEW OF THE POUCH MFR'S TEST RECORDS AND TESTING OF SAMPLE POUCHES TO IDENTIFY WHICH LOTS WERE OUTSIDE OF THE DEFINED ACCEPTANCE PARAMETERS. POUCH PACKETS TESTED AND FOUND TO BE OUTSIDE OF ACCEPTANCE PARAMETERS WERE QUARANTINED AND USING PRODUCTION RECORDS (DEVICE HISTORY RECORDS) ALL POTENTIALLY AFFECTED PRODUCT WAS IDENTIFIED. THIS DATA WAS THEN USED TO CALCULATE THE TOTAL NUMBER OF PRODUCT PACKED IN THE AFFECTED POUCH PACKETS AND RELEASED TO MARKET. AS A RESULT OF THE FINDINGS OF OUR INVESTIGATION, A PRECAUTIONARY RECALL WAS INITIATED IN (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303922 T-FLEX ASPHERIC HQL - INTRAOCULAR LENS HQL RAYNER INTRAOCULAR LENSES LTD. 623T 053E47699

Patients

Seq Age Sex Outcome Treatment
1