ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-10003
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Report Date
- June 16, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/13/2013 WITH THE FOLLOWING FINDINGS: INVESTIGATION CONFIRMED A BLANK DISPLAY SCREEN. THE AUDIBLE TONES WERE FOUND TO BE FUNCTIONING WHEN THE PUMP WAS POWERED ON. UNRELATED TO THE COMPLAINT, THE VIBRATORY FEATURE OF THE PUMP WAS FOUND NOT TO WORKING. ALSO UNRELATED TO THE COMPLAINT, A LEAK TEST REVEALED THAT THE BATTERY COMPARTMENT WAS LEAKING. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE USER GUIDE WARNS THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND/OR THE WATERPROOF FEATURE OF THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN W/MOISTURE) ISSUE AND THERE IS CORROSION IN BATTERY COMPARTMENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312568 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR |