ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-10015
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Report Date
- June 17, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 DATE OF SUBMISSION 09/09/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/15/2013 WITH THE FOLLOWING FINDINGS:DURING TESTING, THE PUMP WAS POWERED ON AND THE DISPLAY SCREEN APPEARED DIM, DISCOLORED, AND DIFFICULT TO READ. WHEN REPLACED WITH A TEST DISPLAY, THE SCREEN FUNCTIONED PROPERLY. UNRELATED TO THE COMPLAINT, ALL OF THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE PRESSES WITH INCREASED FORCE TO ENGAGE. NO DAMAGE WAS OBSERVED TO THE KEYPAD. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THE REPORTER STATED THAT THE DISPLAY FADED IMMEDIATELY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312429 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR |