FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3213897 · Received July 9, 2013

Report

Report Number
2531779-2013-10019
Event Type
Malfunction
Date Received
July 9, 2013
Report Date
June 24, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 09/12/2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/21/2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE FULLY INTACT; NO PEELING WAS OBSERVED. THE COMPLAINT OF INTERMITTENTLY UNRESPONSIVE KEYPAD BUTTONS WAS NOT DUPLICATED DURING TESTING. ALL OF THE KEYPAD BUTTONS RESPONDED APPROPRIATELY AND WERE FUNCTIONAL. THE PUMP WAS OPENED AND REVEALED EVIDENCE OF CONTAMINATION UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE UP, DOWN, AND OK BUTTONS WERE INTERMITTENTLY NONRESPONSIVE. THE REPORTED ISSUE HAS BEEN OCCURRING FOR THE PAST 3 WEEKS. THE PUMP REMAINS ACTIVE AT THIS TIME. A REPLACEMENT PUMP WILL BE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
313837 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 10 YR