ETS FLEX ARTICNG LNR CUTR
Report
- Report Number
- 3005075853-2013-03458
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Date of Event
- June 13, 2013
- Report Date
- June 18, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A 6R45B CARTRIDGE RELOAD PRESENT. THE RETURNED RELOAD WAS PARTIALLY FIRED 1/2 WHICH INDICATES THAT THE DEVICE'S FIRING CYCLE WAS INTERRUPTED. WHEN FIRING THE DEVICE MAKE SURE THAT THE FIRING STROKE IS COMPLETED. DO NOT PARTIALLY FIRE THE DEVICE. FIRE THE DEVICE BY SQUEEZING THE FIRING TRIGGER COMPLETELY UNTIL IT RESTS ON THE CLOSING TRIGGER. ONCE THE FIRING CYCLE HAS BEEN INITIATED, IT MUST BE COMPLETED. IF RE-INITIATION OF FIRING IS RESUMED, THE DEVICE WILL LOCKOUT. FIRING THROUGH THE LOCKOUT MECHANISM WILL BREAK THE DEVICE. PLEASE REFERENCE THE INSTRUCTION FOR USE FOR MORE INFORMATION. THE RETURNED DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT, AND ALL THE STAPLES FORMED AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE.THE CARTRIDGE WAS LOADED INTO THE DEVICE WITHOUT DIFFICULTIES AND DID NOT FALL OUT DURING THE VISUAL AND FUNCTIONAL TESTING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ON WHICH FIRING(S) DID THIS EVENT OCCUR (1ST, 2ND, 8TH, ETC.) HOW MUCH BLOOD WAS LOSS? DID THE PATIENT REQUIRE A BLOOD TRANSFUSION? WHAT COLOR CARTRIDGE WAS BEING USED? WHAT OTHER COLOR CARTRIDGES WERE USED BEFORE AND AFTER THIS EVENT? WAS THE INSTRUMENT FIRED ACROSS OR NEAR AN EXISTING STAPLE LINE OR CLIP? WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? WERE ANY OF THE FORCES HIGHER OR LOWER THAN EXPECTED (CLOSING, FIRING, OR OPENING)? AFTER USE, DID EACH OF THE TRIGGERS AND BUTTONS AUTOMATICALLY RETURN TO THEIR ORIGINAL (PRE-FIRED) POSITIONS, WITHOUT INTERVENTION? WAS THERE ANY DIFFICULTY REMOVING THE DEVICE FROM THE TISSUE? WHAT IS THE CURRENT STATUS OF THE PATIENT?
IT WAS REPORTED THAT DURING A LAPAROSCOPIC APPENDECTOMY PROCEDURE, WHEN STAPLING ACROSS THE APPENDIX AND CECUM, THE STAPLE CARTRIDGE POPPED OUT OF THE DEVICE. THIS CAUSED AN INCOMPLETE STAPLE LINE AND BLEEDING TO OCCUR. AREA WAS RESTAPLED. CASE COMPLETED WITH SAME DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312225 | ETS FLEX ARTICNG LNR CUTR | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | K4CR2Y |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |