FDA Adverse Event Malfunction Summary report: N

ETS FLEX ARTICNG LNR CUTR

MDR report key: 3213845 · Received July 9, 2013

Report

Report Number
3005075853-2013-03458
Event Type
Malfunction
Date Received
July 9, 2013
Date of Event
June 13, 2013
Report Date
June 18, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A 6R45B CARTRIDGE RELOAD PRESENT. THE RETURNED RELOAD WAS PARTIALLY FIRED 1/2 WHICH INDICATES THAT THE DEVICE'S FIRING CYCLE WAS INTERRUPTED. WHEN FIRING THE DEVICE MAKE SURE THAT THE FIRING STROKE IS COMPLETED. DO NOT PARTIALLY FIRE THE DEVICE. FIRE THE DEVICE BY SQUEEZING THE FIRING TRIGGER COMPLETELY UNTIL IT RESTS ON THE CLOSING TRIGGER. ONCE THE FIRING CYCLE HAS BEEN INITIATED, IT MUST BE COMPLETED. IF RE-INITIATION OF FIRING IS RESUMED, THE DEVICE WILL LOCKOUT. FIRING THROUGH THE LOCKOUT MECHANISM WILL BREAK THE DEVICE. PLEASE REFERENCE THE INSTRUCTION FOR USE FOR MORE INFORMATION. THE RETURNED DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT, AND ALL THE STAPLES FORMED AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE.THE CARTRIDGE WAS LOADED INTO THE DEVICE WITHOUT DIFFICULTIES AND DID NOT FALL OUT DURING THE VISUAL AND FUNCTIONAL TESTING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ON WHICH FIRING(S) DID THIS EVENT OCCUR (1ST, 2ND, 8TH, ETC.) HOW MUCH BLOOD WAS LOSS? DID THE PATIENT REQUIRE A BLOOD TRANSFUSION? WHAT COLOR CARTRIDGE WAS BEING USED? WHAT OTHER COLOR CARTRIDGES WERE USED BEFORE AND AFTER THIS EVENT? WAS THE INSTRUMENT FIRED ACROSS OR NEAR AN EXISTING STAPLE LINE OR CLIP? WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? WERE ANY OF THE FORCES HIGHER OR LOWER THAN EXPECTED (CLOSING, FIRING, OR OPENING)? AFTER USE, DID EACH OF THE TRIGGERS AND BUTTONS AUTOMATICALLY RETURN TO THEIR ORIGINAL (PRE-FIRED) POSITIONS, WITHOUT INTERVENTION? WAS THERE ANY DIFFICULTY REMOVING THE DEVICE FROM THE TISSUE? WHAT IS THE CURRENT STATUS OF THE PATIENT?

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC APPENDECTOMY PROCEDURE, WHEN STAPLING ACROSS THE APPENDIX AND CECUM, THE STAPLE CARTRIDGE POPPED OUT OF THE DEVICE. THIS CAUSED AN INCOMPLETE STAPLE LINE AND BLEEDING TO OCCUR. AREA WAS RESTAPLED. CASE COMPLETED WITH SAME DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
312225 ETS FLEX ARTICNG LNR CUTR STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA K4CR2Y

Patients

Seq Age Sex Outcome Treatment
1